SPECIFICATIONS AND TECHNICAL DATA 

Technical data of Ultrasound 

Acoustic frequency: 1MHz ± 10% 

Generator output: 4.0W ± 20% 

Modulation wave shape: 20Hz ± 10% 

Duty factor: 5%,20%, 50%, 80%, 100% 

Treatment time: 10 min 

Actual radiating area(AER): 4.0cm2 ± 20% 

Max intensity 1.0W/cm2 ± 20% 

RBN(MAX): 5.0 

Program: U 

Beam type: Collimated 

Material of applicator: Aluminium 

Technical data of Electronic Stimulator 

Output characteristics: constant voltage(CV) 

Reading resolution: 1V 

Treatment time :  30 min 

Program : E1~E7

Output wave : Monophasic square pulse 

Frequency : 2~150Hz 

Pulse duration : 60-250 μS 

Maximum output : 80 Vpp (500 Ω Load)

Technical data of combination ultrasound and electrical Stimulator 

Output characteristics: Constant Voltage(CV) 

Reading resolution: 1V 

Treatment Time : 10min 

Program : C1 ~ C7

Output wave : Monophase pulse 

Frequency: 2~150 Hz 

Pulse duration: 60~250 μS 

Maximum output : 80 Vpp (500 Ω Load)

Technical data of ULTRA TENS main part 

Supply voltage : 15V 

Current consumption : 10W 

Working current : < 1.0A 

Safety class : Class II, BF-type 

Dimension : 358mm(L) x 64mm (W) x 97mm(H) 

Protection degree : IPX7 for transducer (5mm) 

Technical data of power supply 

Supply voltage : 100V~240V 

Frequency : 50Hz~60Hz 

Power : 18W 

Output voltage : 15V DC 

Output current : 1.0A 

Dimensions : 110mm(L) x 54mm(W) x 33mm(H) 

Environmental conditions 

Operating conditions :  Environment temperature 5~40°C with a relative humidity of    

                                       30%~75% Atmospheric pressure from 700~1060hPa                                                                                                             

Storage conditions :    Environment temperature:-10~50°C  with a relative humidity of 10%~90%

                                       Atmospheric pressure from 700~1060hPa

 

Accessories

ULTRA TENS : 1
Operating instruction : 1
Electrode 50*100mm :  2
Lead for electrical : 2
Adapter 100-240V 50 or 60 Hz : 1
Lead for adapter
Ultrasound transmission gel : 1